NORTH CAROLINA DWI DEFENSEHuntersville • LocustCall / text (704) 880-9388
Evidence
CASE FILE No. G.S. 20-139.1 · BREATH

The Intox EC/IR II, opened up

Every North Carolina breath-test number after a DWI arrest comes out of one machine: the Intoximeters Intox EC/IR II. This case file explains how it measures alcohol, the exact rules the officer must follow, the records that prove or disprove a valid test, and how defense lawyers challenge it in court.

Background: breath-alcohol devices at an NTSB impaired-driving forum exhibit. NTSB, public domain.

Case summary · read before court
The case in one page

One machine, two sensors, a long list of rules

The Intox EC/IR II is a desktop evidential breath analyzer built by Intoximeters, Inc. of St. Louis. North Carolina's Department of Health and Human Services (DHHS) approved it for implied-consent testing in a rule that took effect on November 1, 2007, and in 2020 it added the "Enhanced" version with serial numbers of 10,000 or higher. It is the machine at the jail, the police station or the state's mobile Breath Alcohol Testing (BAT) unit. It is not the handheld screener an officer uses at the roadside.

The name tells you how it works. EC is the electrochemical fuel cell, which measures how much alcohol is in a fixed sample of breath. IR is the infrared system, which watches the breath as you blow, decides when deep-lung air has arrived, and looks for the signature of mouth alcohol. The result prints in grams of alcohol per 210 liters of breath, the same unit North Carolina uses to define the 0.08 offense.

A number from this machine is only admissible if the State proves it was produced the way the law requires: by an analyst with a current DHHS permit for this instrument, after a 15-minute observation period, with at least two sequential samples that agree within 0.02, on an instrument that received preventive maintenance in the last four months. Each of those requirements leaves a paper trail, and the paper trail is where most breath cases are won or lost.

Exhibit 1Front view of an Intox EC/IR II evidential breath test instrument with breath hose, display, keyboard and printer
The Intox EC/IR II, as pictured in the West Virginia State Police Intox EC/IR II Operator's Manual, Rev. 2 (June 6, 2023), p. 4. Shown for identification; not a photo of any North Carolina case instrument.
15minutes of observation before each breath sample (10A NCAC 41B .0101(6))
30minutes, at most, that testing waits for your lawyer or witness (G.S. 20-16.2(a)(6))
0.02maximum gap between two consecutive samples; the lower one counts (G.S. 20-139.1(b3))
4months, at most, between preventive maintenance checks (10A NCAC 41B .0323)
.000–.440measuring range in g/210 L, per the manufacturer's specification sheet
35minimum course hours for an analyst's initial DHHS permit (10A NCAC 41B .0301)
Evidence board

Everyone and everything in this file

Tap a card to open it. The red string runs in the order a breath test actually happens, from the machine on the counter to the judge who decides whether its number comes in.

THE INSTRUMENTIntoximeters Intox EC/IR II, a dual-sensor benchtop analyzer.
Manufacturer brochure: about 19 x 14.5 x 7 inches, 18.75 pounds, range .000 to .440 g/210 L, radio-frequency interference detection, and an internal thermal printer option. It meets the NHTSA and OIML model specifications for evidential breath testers.
THE FUEL CELL (EC)Measures the alcohol.
A porous disk coated with platinum. Alcohol in a fixed volume of breath, about 1 milliliter in the RCMP's description, is oxidized on the surface and produces an electrical current proportional to the alcohol. Exhibit A
THE INFRARED BENCH (IR)Watches the breath.
The RCMP's training text says it monitors three wavelengths: two for ethanol and one for carbon dioxide. It tracks the breath in real time to confirm deep-lung air and flag mouth alcohol.
THE CHEMICAL ANALYSTUsually the arresting officer or another officer.
Must hold a current DHHS permit for the Intox EC/IR II on the test date. DHHS must post permits online, and courts take judicial notice of them under G.S. 20-139.1(b6).
THE REGULATORNC DHHS Forensic Tests for Alcohol Branch.
Writes the rules in 10A NCAC 41B, trains and permits analysts, performs preventive maintenance, runs seven BAT mobile units, and posts monthly maintenance records. Exhibit C
THE RULES41B .0101, .0322, .0323.
.0101(6) defines the 15-minute observation period. .0322 lists the eight operating steps. .0323 sets four-month preventive maintenance. Exhibit B
THE TICKETThe printed test record.
Serial number, date and time, air blanks, the accuracy check against a known 0.08 standard, each subject sample, the reported result, the analyst and permit number. Read it line by line.
THE COURTSWhere the number is tested.
State v. Vaughn (2024): no foundation that the two samples agreed within 0.02, new trial. State v. Forney (2024): gum during observation was error. Exhibit D
Chronology

How North Carolina got here

  1. 1964The state's breath-alcohol program, today the DHHS Forensic Tests for Alcohol Branch, begins training North Carolina officers in breath testing.
  2. 1984State v. Shuping, 312 N.C. 421: the Supreme Court of North Carolina holds that the instrument's accuracy-check tolerance protects the defendant rather than creating a margin of error in the subject's result.
  3. Dec. 1, 2006Session Law 2006-253 requires "at least duplicate sequential breath samples" and makes a valid result sufficient evidence of alcohol concentration.
  4. Nov. 1, 200710A NCAC 41B .0322, the Intox EC/IR II operating rule, takes effect.
  5. Dec. 17, 2007NHTSA adds the "EC/IR II (Enhanced with serial number 10,000 or higher)" to its Conforming Products List, 72 Fed. Reg. 71480.
  6. 2008The last Intoxilyzer 5000 maintenance histories DHHS posts. EC/IR II records continue to the present.
  7. Apr. 1, 2020North Carolina amends .0322 and .0323 to cover the Enhanced model, 34:11 N.C. Reg. 908–909.
  8. Jan. 16, 2024State v. Forney: gum in the mouth during observation required a new observation period; the error was found harmless on that record.
  9. Dec. 3, 2024State v. Vaughn: an EC/IR II result admitted without proof that the two samples agreed within 0.02 earns a new trial.
  10. 2025–2026Virginia, which adopted the EC/IR II in 2008, buys 220 Dräger Alcotest 9510 instruments to replace it, with deployment expected to begin in late 2026.
From the defense table

Five questions that decide most breath cases

After years on both sides of the breath room, first as a Charlotte-Mecklenburg police officer who made several hundred DWI arrests and now as defense counsel, Chris Adkins looks at every Intox EC/IR II result through the same five questions.

  1. Was the stop and arrest lawful? A perfect breath test does not survive an unlawful stop or an arrest without probable cause.
  2. Were you advised of your rights, orally and in writing, and given your 30 minutes? A witness who arrived in time and was turned away can sink the result. See State v. Hatley and State v. Myers in Exhibit D.
  3. Was there a real 15-minute observation before each sample? Nothing to eat, drink or smoke, no vomiting or regurgitation. Gum, tobacco, a burp or a trip out of the room all deserve a hard look.
  4. Do the samples meet G.S. 20-139.1(b3)? Two consecutive results within 0.02, the lower one reported, and proof of both in the record.
  5. Were the analyst and the machine certified? A current permit for this instrument type, and preventive maintenance inside the four-month window.

From the officer's side

The breath room is a small space with a lot happening in a few minutes: a rights form, phone calls, a witness who may or may not be in the lobby, data entry, and a person who needs watching. The rule lets the analyst observe while operating the instrument, and good analysts do both. When the times on the rights form, the observation start and the test record don't line up, that gap is usually where the real story is.

The manual question

Where is the North Carolina Intox EC/IR II manual?

We looked hard. As of November 2026, North Carolina does not publish its Intox EC/IR II operator manual or course materials online, and we found no publicly released copy of the manufacturer's own operator or service manual. DHHS's Forensic Tests for Alcohol Branch lists its Intox EC/IR II Operator and Recertification courses, but the course materials themselves are not posted.

What is public: North Carolina's operating and maintenance rules, the monthly maintenance records for every instrument, permit histories, and several other states' government-written EC/IR II manuals that describe the same instrument. Exhibit C collects all of them, with what each one is good for and where its settings differ from North Carolina's.

Open the manual library →

Read the label first

Another state's manual describes the same hardware but that state's own settings, such as a 20-minute waiting period in West Virginia, or a different accuracy-check window. In a North Carolina courtroom, North Carolina's rules control.

Exhibit E preview

Is North Carolina replacing the Intox EC/IR II?

Not that we can find. We searched DHHS rulemaking, the Forensic Tests for Alcohol Branch pages and General Assembly bills through November 2026 and found no announced plan, budget item or procurement to replace North Carolina's EC/IR II fleet. The 2020 rule change added the Enhanced model; it did not retire the original.

Next door is a different story. Virginia, which approved the EC/IR II in 2008, issued a notice of award for the Dräger Alcotest 9510 on March 7, 2025, completed the purchase of 220 instruments, and expects to begin evidential testing with them in late 2026. Intoximeters has not released a successor model; the federal list of approved evidential instruments still shows the EC/IR II, the Enhanced EC/IR II and the transportable EC/IR II.t.

Read the full story →

Test yourself

Intox EC/IR II quiz

Which part of the Intox EC/IR II produces the reported alcohol number?

The fuel cell quantifies the alcohol. Infrared monitors the breath, decides when deep-lung air has arrived and flags mouth alcohol.

Your two samples read 0.10 and 0.09. What is reported?

They are within 0.02, and G.S. 20-139.1(b3) says only the lower of the two consecutive results can be used.

How long must the analyst observe you before each sample?

10A NCAC 41B .0101(6). The 30-minute figure is the separate window for your lawyer or witness. West Virginia's manual uses 20 minutes; North Carolina does not.

You give one good sample, then refuse the second. What happens?

G.S. 20-139.1(b3) treats refusing a second or later sample as a refusal under G.S. 20-16.2(c) and makes the first result admissible.

Has North Carolina announced a replacement for the EC/IR II?

Virginia is switching to the Alcotest 9510. We found no North Carolina plan.

Questions people ask

Intox EC/IR II: frequently asked questions

What breath machine does North Carolina use after a DWI arrest?

The Intoximeters Intox EC/IR II and the Intox EC/IR II Enhanced (serial number 10,000 or higher), approved under 10A NCAC 41B .0322. The rule took effect November 1, 2007, and was amended to cover the Enhanced model effective April 1, 2020.

Is the Intox EC/IR II the same as the roadside breathalyzer?

No. Roadside devices are alcohol screening tests under G.S. 20-16.3, and generally only the fact that alcohol was or was not detected can be used, not the number. The EC/IR II is the evidential instrument whose numeric result can prove the 0.08 offense.

How does the Intox EC/IR II measure alcohol?

An infrared system monitors the breath for ethanol and carbon dioxide as you exhale. When it detects deep-lung air, a small fixed volume is drawn into an electrochemical fuel cell, where alcohol is oxidized on a platinum surface and produces an electrical current proportional to the alcohol in the sample.

Why do I have to blow twice?

G.S. 20-139.1(b3) requires at least duplicate sequential breath samples. Two consecutive results must agree within 0.02, and only the lower one may be used. If they differ by more, the rule calls for a third or fourth sample.

Can burping, vomiting, gum or tobacco affect the test?

They can introduce mouth alcohol or other material, which is why the 15-minute observation period forbids eating, drinking, smoking, vomiting and regurgitation. In State v. Forney (2024), the Court of Appeals held that gum found in the mouth after the first sample required a new observation period.

Where can I find the North Carolina Intox EC/IR II manual?

North Carolina does not post its operator manual or course materials online. The rules, monthly maintenance records and permit histories are public on the DHHS Forensic Tests for Alcohol Branch website, and West Virginia, Virginia and Arkansas publish their own state-written EC/IR II manuals. Exhibit C links all of them.

Is North Carolina getting a new breath-test machine?

We found no announced plan as of November 2026. Virginia has bought 220 Dräger Alcotest 9510 instruments to replace its EC/IR II fleet, with deployment expected to start in late 2026.

Can an Intox EC/IR II result be thrown out?

Yes, when the State cannot show the test followed the DHHS rules and was given by a permitted analyst, when a witness or rights violation occurred, when maintenance was out of date and the defense objects and proves it, or when the stop or arrest was unlawful. Each case turns on its own records.

Keep reading

Related guides

Chris Adkins, North Carolina DWI defense attorney and former Charlotte-Mecklenburg police officer
By Chris Adkins and Chris Purkey, Adkins & Purkey, PLLC. Chris Adkins is a former Charlotte-Mecklenburg police officer who made several hundred DWI arrests before becoming a defense attorney. Together they have defended hundreds of DWI clients and co-wrote Charged with DWI in North Carolina? Reviewed November 2026.

Blew into an Intox EC/IR II? Let's read the record together.

We review the test ticket, the rights form, the observation timeline, the analyst's permit and the instrument's maintenance history against North Carolina's rules. Offices in Huntersville (Lake Norman) and Locust (Stanly County), serving Mecklenburg, Iredell, Cabarrus, Stanly and courts across North Carolina.

General North Carolina legal information current as of November 2026, not legal advice for any specific case. Reading this page does not create an attorney-client relationship. Attorney advertising. Intox EC/IR II is a product name of Intoximeters, Inc.; this site is not affiliated with the manufacturer or with NC DHHS.

AI DWI guide · Powered by Google Gemini

Ask DWI Dave a question

I’m DWI Dave, LKN DWI’s robot guide. Type any North Carolina DWI question in plain English and I’ll answer in seconds from our library, with links to the pages that cover it.

General information, not legal advice. Please don't type names, case numbers or other confidential details.

ADKINS & PURKEY, PLLC

Your next move starts here.

Talk with the firm about your charge, your license, and the path ahead.

Call (704) 880-9388Text the firm