EXHIBIT A · THE SCIENCE

Inside the machine

The Intox EC/IR II uses two different kinds of physics on one breath. Infrared light watches the breath as it flows; a platinum fuel cell turns the alcohol in a small sample into electricity. Here is how each one works, what each one can and cannot tell you, and why it matters in a North Carolina courtroom.

Background: a breath alcohol reading during a 2013 safe-driving study. U.S. Air Force photo by Senior Airman Natasha Stannard, public domain.

Exhibit A · laboratory analysis
Start here

What the machine is actually measuring

North Carolina law defines "alcohol concentration" two ways: grams of alcohol per 100 milliliters of blood, or grams of alcohol per 210 liters of breath (G.S. 20-4.01(1b)). The Intox EC/IR II reports the second one. When the ticket says 0.09, it means 0.09 grams of ethanol for every 210 liters of breath, a concentration, not a volume you had to blow. The instrument analyzes a sample of roughly a milliliter and scales the result to the legal unit.

That distinction matters more than it sounds. Because the offense itself is written in breath units, a valid breath result proves the "0.08 or more" element directly, without converting anything to blood. The science still matters, though, because every step between your lungs and the printed number is a place where something can go right or wrong.

The two sensors in one sentence each

Infrared (IR): decides when to sample and whether the breath looks like clean deep-lung air. Electrochemical fuel cell (EC): decides how much alcohol is in the sample that gets reported. The instrument does not average two independent alcohol readings; each sensor has its own job.

Diagram

The path of one breath

Government training materials for the EC/IR II describe the same basic route. The Royal Canadian Mounted Police's Intox EC/IR II resource text (2018) describes a temperature-controlled breath tube, an infrared sample chamber of about 9 milliliters, and a fuel cell that draws a fixed sample of about 1 milliliter. Virginia's Department of Forensic Science describes a tungsten infrared source and three filters, two for ethanol and one for carbon dioxide.

Simplified diagram of breath flow through an Intox EC/IR IIBreath enters a mouthpiece and heated breath tube, passes through an infrared sample chamber with a light source on one side and filters and detectors on the other, then a valve sends a fixed sample to the fuel cell while the rest exhausts. Mouth Heated breath tubeRCMP: 40 °C ± 1 °C Infrared sample chamber (about 9 mL) IR source Filters + detectors 2 ethanol channels1 CO₂ channel Valve Fuel cell~1 mL sample Exhaust Simplified illustration, not to scale. Based on RCMP/Alberta (2018) and Virginia DFS (2025 draft) training descriptions; component details can vary by configuration.
Original illustration by LKN DWI. It shows the order of events, not engineering drawings.
Sensor one

Infrared: every molecule has a fingerprint

Molecules vibrate, and each kind of chemical bond absorbs infrared light at its own characteristic wavelengths. Shine infrared light through a gas, measure how much comes out the other side at a wavelength the target molecule absorbs, and you can tell how much of that molecule is in the beam. The relationship is the Beer–Lambert law: absorbance rises with the concentration of the absorbing gas and with the length of the light path.

The EC/IR II uses that principle on the moving breath stream. According to the RCMP's training text, its infrared detector monitors three wavelengths: two for ethanol and one for carbon dioxide. The manufacturer says the infrared system "monitor[s] the breath sample for alcohol and carbon dioxide."

Why carbon dioxide? Human breath carries CO₂ from the lungs; a gas standard and room air do not. Watching CO₂ alongside ethanol helps the instrument tell a real exhalation from something else and follow the breath as it changes from mouth air to deep-lung air.

Exhibit A-1Mid-infrared absorption spectra of carbon dioxide, water, carbon monoxide, methane and other gases, with the strong CO2 band near 4.3 micrometers
Every gas absorbs infrared light in its own bands; carbon dioxide's strong band near 4.3 µm stands out. Chart: Daniel Popa and Florin Udrea, Sensors (2019), HITRAN data, CC BY 4.0, via Wikimedia Commons. Ethanol is not plotted here.

What we could not verify

Many websites state the exact infrared wavelengths the EC/IR II uses. We could not confirm them in a manufacturer or government document, so we do not repeat them here. If wavelengths matter in a case, the instrument's documentation or a qualified expert is the right source.

Interactive lab

Watch a breath the way the infrared detector does

As you blow, the first air out is from your mouth and upper airway. Then comes air from deeper in the lungs, where breath has been in contact with blood in the alveoli. A good sample rises and levels off. Mouth alcohol does the opposite: it spikes early, then falls as cleaner lung air pushes it out. Pick a scenario below. The curves are teaching illustrations, not data from any instrument.

Breath profile simulator

That falling-slope pattern is why the RCMP text describes mouth alcohol as producing "a large decrease in transmitted energy at the ethyl alcohol wavelengths" early in the blow. When the instrument sees it, it is supposed to stop and report a problem rather than print a number. The human safeguard is the 15-minute observation period, which exists so mouth alcohol has time to dissipate and so the analyst can see a burp, a drink or something in the mouth.

Two practical limits: a detection feature only catches what its programmed criteria are built to catch, and the absence of a warning on a ticket is evidence the criteria were not triggered, not proof that nothing was in the mouth. Conditions like GERD, dentures and recent dental work are covered in our article on mouth alcohol, GERD and diabetes.

Exhibit A-2Handheld breath alcohol tester labeled Fuel Cell Technology showing a reading on its display
A handheld tester labeled "Fuel Cell Technology." The EC/IR II uses the same basic electrochemical sensor inside a larger evidential instrument. Photo: Ditri, CC BY-SA 4.0, via Wikimedia Commons.
Sensor two

The fuel cell: turning alcohol into electricity

A breath-alcohol fuel cell is a small sandwich: a porous, chemically inert disk coated on both sides with finely divided platinum, soaked with an acidic electrolyte, with wires to each face. The RCMP's training text describes exactly that construction for the EC/IR II.

When a fixed volume of breath is drawn across the cell, ethanol is oxidized at one face. In simplified form:

Anode (alcohol side)CH₃CH₂OH → CH₃CHO + 2H⁺ + 2e⁻
Cathode (oxygen side)½O₂ + 2H⁺ + 2e⁻ → H₂O

The electrons released by the alcohol travel through the external circuit, and that flow is the signal. More ethanol molecules in the sample means more charge. The instrument's processor converts the measured electrical response into grams per 210 liters using its calibration.

Fuel cells are popular for evidential testing because they are quite specific to alcohols. The manufacturer says the EC/IR II "does not respond to acetone," a compound that can be elevated in the breath of people with diabetes or on very low-calorie diets. Other alcohols such as methanol and isopropanol do react, and the RCMP text describes the instrument comparing reaction profiles to flag an interfering substance; Virginia's manual lists an "INTERFERENT DETECTED" message.

Quality control

Blanks, standards and the 0.08 accuracy check

A breath test sequence is more than your two blows. Around each subject sample the instrument runs checks that are printed on the ticket:

Room air is drawn through the system and should read zero. A blank shows whether alcohol or another substance is lingering in the instrument or the room before a sample is analyzed.

The instrument analyzes a reference sample with a known concentration equivalent to 0.08. Under 10A NCAC 41B .0101, a result at the expected 0.08 or 0.01 below it is accepted. In State v. Shuping, 312 N.C. 421 (1984), the Supreme Court of North Carolina treated that kind of tolerance as a safeguard favoring the defendant, not as a margin of error to subtract from the subject's result.

An ethanol gas canister is a compressed cylinder of alcohol in an inert gas at an accurately known concentration. North Carolina's maintenance rule requires a technician to verify the canister shows at least 51 pounds per square inch of pressure and that canisters are changed before their expiration date.

A simulator is a jar of water-alcohol solution held at a constant temperature, producing an alcohol-air mixture of known concentration above the liquid. NC requires the thermometer to show 34 °C ± 0.2 °C and the solution to be changed every four months or after 125 tests, whichever comes first (10A NCAC 41B .0323). The 2007 federal listing of the Enhanced EC/IR II mentions hardware for recirculating a wet bath simulator.

Why this matters in court

A passing accuracy check tells you the instrument read a known standard correctly at that moment. It does not prove the observation period happened, that the right person blew, or that a later sample was clean. Each of those needs its own evidence. Exhibit C explains how to match the reference lot, expiration date and maintenance record to the test.

Interactive lab

The 2,100-to-1 partition ratio

Why 210 liters? Breath testing grew up on the assumption that, at equilibrium in the lungs, the alcohol in 2,100 milliliters of deep-lung air roughly equals the alcohol in 1 milliliter of blood. That convention is built into the unit: 0.08 g per 210 liters of breath was chosen to line up with 0.08 g per 100 milliliters of blood.

Real people are not all exactly 2,100 to 1. Research shows the ratio varies between individuals and can change during the absorption and elimination of alcohol. Move the slider to see what a different personal ratio would mean physiologically.

Breath-to-blood converter (teaching tool)

In North Carolina this is mostly an impairment and expert-testimony issue, not a per se defense, because the statute measures alcohol concentration in breath units directly. Our deeper article explains the research and how it is used: the 2100:1 partition ratio.

Specification sheet

The Intox EC/IR II by the numbers

Sources: Intoximeters law-enforcement brochure (ECIRII-LE, rev. 11/24/18); NHTSA Conforming Products List, 72 Fed. Reg. 71480 (Dec. 17, 2007) and 82 Fed. Reg. 50942 (Nov. 2, 2017); RCMP/Alberta Intox EC/IR II Resource Reading Material (2018). The Alberta flow and volume settings are that jurisdiction's configuration; North Carolina's settings are set by DHHS and are not published in the rules.
FeatureWhat the source saysSource
ManufacturerIntoximeters, Inc., St. Louis, MissouriManufacturer
Alcohol sensorElectrochemical fuel cellManufacturer
Breath monitoringInfrared detection of alcohol and carbon dioxide; three wavelengths (two ethanol, one CO₂)Manufacturer; RCMP
Measuring range.000 to .440 g/210 LManufacturer
Standards metNHTSA and OIML model specifications for evidential breath testersManufacturer
InterferenceRadio-frequency interference detection; does not respond to acetoneManufacturer
Size and weight19 × 14.5 × 7 inches; 18.75 poundsManufacturer
Display and printer256 × 32 graphic vacuum fluorescent display; optional internal thermal printerManufacturer
ConnectionsRS232, parallel, USB, PS/2 and EthernetManufacturer
Enhanced modelAdded test memory, hardware to recirculate a wet bath simulator, enhanced EMC and RFI immunity (serial numbers 10,000 and up)72 Fed. Reg. 71480
Sample volumesIR chamber about 9 mL; fuel cell sample about 1 mL; breath tube 40 °C ± 1 °CRCMP
Breath acceptance (Alberta)Minimum flow 12 L/min; minimum volume 1.5 LRCMP (jurisdiction setting)
Federal listingEC/IR, EC/IR II, EC/IR II Enhanced and the transportable EC/IR II.t appear on the federal list of evidential breath devices82 Fed. Reg. 50942; U.S. DOT list (Jan. 2025)

Questions worth taking to a qualified expert

  • What software version and configuration was this instrument running on the test date, and what were its breath-acceptance criteria?
  • Do the instrument's stored records show any aborted attempts, warnings or interferent or mouth-alcohol messages before the reported samples?
  • Was the reference a dry gas canister or a wet bath simulator, and do the lot, expiration and maintenance records match the ticket?
  • Does anything in this person's medical history (reflux, dentures, a respiratory condition) bear on the observation period or the ability to give a sample?
  • For a number near 0.08, what does the full sequence (blanks, check, both samples) show, and does the record prove the two samples agreed within 0.02?

Exhibit A: frequently asked questions

Does the Intox EC/IR II average the infrared and fuel cell readings?

No. Infrared monitors the breath and decides when deep-lung air has arrived; the fuel cell measures the alcohol in the sample that is reported.

Do I really blow 210 liters of air?

No. Grams per 210 liters is a concentration unit. The instrument analyzes a small fixed sample, about a milliliter according to the RCMP's description, and reports the concentration in the legal unit.

Can diabetes cause a false reading on the Intox EC/IR II?

The manufacturer states the instrument does not respond to acetone, the compound most associated with diabetic ketosis. Other medical conditions, such as reflux, raise mouth-alcohol and observation questions instead.

What happens if the instrument detects mouth alcohol or an interfering substance?

It is designed to stop and display a message rather than report a result. The full sequence, including aborted attempts, can be important evidence.

Sources for Exhibit A

Chris Adkins, North Carolina DWI defense attorney
By Chris Adkins and Chris Purkey, Adkins & Purkey, PLLC. Former Charlotte-Mecklenburg police officer turned DWI defense attorney, and co-authors of Charged with DWI in North Carolina? For the scholarly companion, see The Lake Norman Brief's instrument-science chapter. Reviewed November 2026.

Questions about your breath result?

We review the ticket, the instrument's records and the science with you. Huntersville and Locust offices, serving courts across North Carolina.

General North Carolina legal and scientific information current as of November 2026, not legal advice. Diagrams and simulations are original teaching illustrations, not instrument data. Intox EC/IR II is a product name of Intoximeters, Inc.; this site is not affiliated with the manufacturer or NC DHHS. Attorney advertising.

ADKINS & PURKEY, PLLC

Your next move starts here.

Talk with the firm about your charge, your license, and the path ahead.

Call (704) 880-9388Text the firm