The Intox EC/IR II uses two different kinds of physics on one breath. Infrared light watches the breath as it flows; a platinum fuel cell turns the alcohol in a small sample into electricity. Here is how each one works, what each one can and cannot tell you, and why it matters in a North Carolina courtroom.
Background: a breath alcohol reading during a 2013 safe-driving study. U.S. Air Force photo by Senior Airman Natasha Stannard, public domain.
What the machine is actually measuring
North Carolina law defines "alcohol concentration" two ways: grams of alcohol per 100 milliliters of blood, or grams of alcohol per 210 liters of breath (G.S. 20-4.01(1b)). The Intox EC/IR II reports the second one. When the ticket says 0.09, it means 0.09 grams of ethanol for every 210 liters of breath, a concentration, not a volume you had to blow. The instrument analyzes a sample of roughly a milliliter and scales the result to the legal unit.
That distinction matters more than it sounds. Because the offense itself is written in breath units, a valid breath result proves the "0.08 or more" element directly, without converting anything to blood. The science still matters, though, because every step between your lungs and the printed number is a place where something can go right or wrong.
The two sensors in one sentence each
Infrared (IR): decides when to sample and whether the breath looks like clean deep-lung air. Electrochemical fuel cell (EC): decides how much alcohol is in the sample that gets reported. The instrument does not average two independent alcohol readings; each sensor has its own job.
The path of one breath
Government training materials for the EC/IR II describe the same basic route. The Royal Canadian Mounted Police's Intox EC/IR II resource text (2018) describes a temperature-controlled breath tube, an infrared sample chamber of about 9 milliliters, and a fuel cell that draws a fixed sample of about 1 milliliter. Virginia's Department of Forensic Science describes a tungsten infrared source and three filters, two for ethanol and one for carbon dioxide.
Infrared: every molecule has a fingerprint
Molecules vibrate, and each kind of chemical bond absorbs infrared light at its own characteristic wavelengths. Shine infrared light through a gas, measure how much comes out the other side at a wavelength the target molecule absorbs, and you can tell how much of that molecule is in the beam. The relationship is the Beer–Lambert law: absorbance rises with the concentration of the absorbing gas and with the length of the light path.
The EC/IR II uses that principle on the moving breath stream. According to the RCMP's training text, its infrared detector monitors three wavelengths: two for ethanol and one for carbon dioxide. The manufacturer says the infrared system "monitor[s] the breath sample for alcohol and carbon dioxide."
Why carbon dioxide? Human breath carries CO₂ from the lungs; a gas standard and room air do not. Watching CO₂ alongside ethanol helps the instrument tell a real exhalation from something else and follow the breath as it changes from mouth air to deep-lung air.

What we could not verify
Many websites state the exact infrared wavelengths the EC/IR II uses. We could not confirm them in a manufacturer or government document, so we do not repeat them here. If wavelengths matter in a case, the instrument's documentation or a qualified expert is the right source.
Watch a breath the way the infrared detector does
As you blow, the first air out is from your mouth and upper airway. Then comes air from deeper in the lungs, where breath has been in contact with blood in the alveoli. A good sample rises and levels off. Mouth alcohol does the opposite: it spikes early, then falls as cleaner lung air pushes it out. Pick a scenario below. The curves are teaching illustrations, not data from any instrument.
Breath profile simulator
That falling-slope pattern is why the RCMP text describes mouth alcohol as producing "a large decrease in transmitted energy at the ethyl alcohol wavelengths" early in the blow. When the instrument sees it, it is supposed to stop and report a problem rather than print a number. The human safeguard is the 15-minute observation period, which exists so mouth alcohol has time to dissipate and so the analyst can see a burp, a drink or something in the mouth.
Two practical limits: a detection feature only catches what its programmed criteria are built to catch, and the absence of a warning on a ticket is evidence the criteria were not triggered, not proof that nothing was in the mouth. Conditions like GERD, dentures and recent dental work are covered in our article on mouth alcohol, GERD and diabetes.

The fuel cell: turning alcohol into electricity
A breath-alcohol fuel cell is a small sandwich: a porous, chemically inert disk coated on both sides with finely divided platinum, soaked with an acidic electrolyte, with wires to each face. The RCMP's training text describes exactly that construction for the EC/IR II.
When a fixed volume of breath is drawn across the cell, ethanol is oxidized at one face. In simplified form:
Cathode (oxygen side)½O₂ + 2H⁺ + 2e⁻ → H₂O
The electrons released by the alcohol travel through the external circuit, and that flow is the signal. More ethanol molecules in the sample means more charge. The instrument's processor converts the measured electrical response into grams per 210 liters using its calibration.
Fuel cells are popular for evidential testing because they are quite specific to alcohols. The manufacturer says the EC/IR II "does not respond to acetone," a compound that can be elevated in the breath of people with diabetes or on very low-calorie diets. Other alcohols such as methanol and isopropanol do react, and the RCMP text describes the instrument comparing reaction profiles to flag an interfering substance; Virginia's manual lists an "INTERFERENT DETECTED" message.
Blanks, standards and the 0.08 accuracy check
A breath test sequence is more than your two blows. Around each subject sample the instrument runs checks that are printed on the ticket:
Room air is drawn through the system and should read zero. A blank shows whether alcohol or another substance is lingering in the instrument or the room before a sample is analyzed.
The instrument analyzes a reference sample with a known concentration equivalent to 0.08. Under 10A NCAC 41B .0101, a result at the expected 0.08 or 0.01 below it is accepted. In State v. Shuping, 312 N.C. 421 (1984), the Supreme Court of North Carolina treated that kind of tolerance as a safeguard favoring the defendant, not as a margin of error to subtract from the subject's result.
An ethanol gas canister is a compressed cylinder of alcohol in an inert gas at an accurately known concentration. North Carolina's maintenance rule requires a technician to verify the canister shows at least 51 pounds per square inch of pressure and that canisters are changed before their expiration date.
A simulator is a jar of water-alcohol solution held at a constant temperature, producing an alcohol-air mixture of known concentration above the liquid. NC requires the thermometer to show 34 °C ± 0.2 °C and the solution to be changed every four months or after 125 tests, whichever comes first (10A NCAC 41B .0323). The 2007 federal listing of the Enhanced EC/IR II mentions hardware for recirculating a wet bath simulator.
Why this matters in court
A passing accuracy check tells you the instrument read a known standard correctly at that moment. It does not prove the observation period happened, that the right person blew, or that a later sample was clean. Each of those needs its own evidence. Exhibit C explains how to match the reference lot, expiration date and maintenance record to the test.
The 2,100-to-1 partition ratio
Why 210 liters? Breath testing grew up on the assumption that, at equilibrium in the lungs, the alcohol in 2,100 milliliters of deep-lung air roughly equals the alcohol in 1 milliliter of blood. That convention is built into the unit: 0.08 g per 210 liters of breath was chosen to line up with 0.08 g per 100 milliliters of blood.
Real people are not all exactly 2,100 to 1. Research shows the ratio varies between individuals and can change during the absorption and elimination of alcohol. Move the slider to see what a different personal ratio would mean physiologically.
Breath-to-blood converter (teaching tool)
In North Carolina this is mostly an impairment and expert-testimony issue, not a per se defense, because the statute measures alcohol concentration in breath units directly. Our deeper article explains the research and how it is used: the 2100:1 partition ratio.
The Intox EC/IR II by the numbers
| Feature | What the source says | Source |
|---|---|---|
| Manufacturer | Intoximeters, Inc., St. Louis, Missouri | Manufacturer |
| Alcohol sensor | Electrochemical fuel cell | Manufacturer |
| Breath monitoring | Infrared detection of alcohol and carbon dioxide; three wavelengths (two ethanol, one CO₂) | Manufacturer; RCMP |
| Measuring range | .000 to .440 g/210 L | Manufacturer |
| Standards met | NHTSA and OIML model specifications for evidential breath testers | Manufacturer |
| Interference | Radio-frequency interference detection; does not respond to acetone | Manufacturer |
| Size and weight | 19 × 14.5 × 7 inches; 18.75 pounds | Manufacturer |
| Display and printer | 256 × 32 graphic vacuum fluorescent display; optional internal thermal printer | Manufacturer |
| Connections | RS232, parallel, USB, PS/2 and Ethernet | Manufacturer |
| Enhanced model | Added test memory, hardware to recirculate a wet bath simulator, enhanced EMC and RFI immunity (serial numbers 10,000 and up) | 72 Fed. Reg. 71480 |
| Sample volumes | IR chamber about 9 mL; fuel cell sample about 1 mL; breath tube 40 °C ± 1 °C | RCMP |
| Breath acceptance (Alberta) | Minimum flow 12 L/min; minimum volume 1.5 L | RCMP (jurisdiction setting) |
| Federal listing | EC/IR, EC/IR II, EC/IR II Enhanced and the transportable EC/IR II.t appear on the federal list of evidential breath devices | 82 Fed. Reg. 50942; U.S. DOT list (Jan. 2025) |
Questions worth taking to a qualified expert
- What software version and configuration was this instrument running on the test date, and what were its breath-acceptance criteria?
- Do the instrument's stored records show any aborted attempts, warnings or interferent or mouth-alcohol messages before the reported samples?
- Was the reference a dry gas canister or a wet bath simulator, and do the lot, expiration and maintenance records match the ticket?
- Does anything in this person's medical history (reflux, dentures, a respiratory condition) bear on the observation period or the ability to give a sample?
- For a number near 0.08, what does the full sequence (blanks, check, both samples) show, and does the record prove the two samples agreed within 0.02?
Exhibit A: frequently asked questions
Does the Intox EC/IR II average the infrared and fuel cell readings?
No. Infrared monitors the breath and decides when deep-lung air has arrived; the fuel cell measures the alcohol in the sample that is reported.
Do I really blow 210 liters of air?
No. Grams per 210 liters is a concentration unit. The instrument analyzes a small fixed sample, about a milliliter according to the RCMP's description, and reports the concentration in the legal unit.
Can diabetes cause a false reading on the Intox EC/IR II?
The manufacturer states the instrument does not respond to acetone, the compound most associated with diabetic ketosis. Other medical conditions, such as reflux, raise mouth-alcohol and observation questions instead.
What happens if the instrument detects mouth alcohol or an interfering substance?
It is designed to stop and display a message rather than report a result. The full sequence, including aborted attempts, can be important evidence.
Sources for Exhibit A
- StatuteG.S. 20-4.01(1b), definition of alcohol concentration
- Rule10A NCAC 41B .0101 definitions; .0323 preventive maintenance
- ManufacturerIntoximeters, Intox EC/IR II law-enforcement brochure and product page
- TrainingRCMP / Alberta, Intox EC/IR II Resource Reading Material (2018) (Crown copyright; linked, not reproduced)
- TrainingVirginia DFS, draft Intox EC/IR II manual (2025) and 2008 instructional manual
- Federal72 Fed. Reg. 71480 (Dec. 17, 2007); 82 Fed. Reg. 50942 (Nov. 2, 2017); U.S. DOT approved evidential breath devices
- CaseState v. Shuping, 312 N.C. 421, 323 S.E.2d 350 (1984)
Questions about your breath result?
We review the ticket, the instrument's records and the science with you. Huntersville and Locust offices, serving courts across North Carolina.
General North Carolina legal and scientific information current as of November 2026, not legal advice. Diagrams and simulations are original teaching illustrations, not instrument data. Intox EC/IR II is a product name of Intoximeters, Inc.; this site is not affiliated with the manufacturer or NC DHHS. Attorney advertising.